Start Up Procedure
- Remove Packaging: Ensure all internal and external packaging materials, including tape, plastic bags, and foam, are removed from the cabinet.
- Power On: Place the cabinet in its designated location, with water and drain connections set up.
- Connect Power: Insert the supplied lead into the IEC socket on the back of the cabinet, then plug the 3-pin connector into a 15A general-purpose outlet.
- Warm-Up: The controller will initiate a warm-up cycle, after which it will display the security screensaver (SOV mode).
Factory Settings:
- On the first start-up, the temperature is preset to 25°C, and the humidity function is turned off by default.
Shut Down Procedure
-
Power Off: Turning off the mains power will trigger a power fail alarm due to the battery-backed alarm system.
-
Controller Shutdown: Press and hold the red button while switching off the main power to power down the controller fully.
To access the full instructions for the Star X controller, please refer to the Star-X manual.
Manual Reset Safety Thermostat
The over-temperature safety thermostat is not operator adjustable. It will electrically isolate the heating elements in the event of an over-temperature condition. This safety feature prevents overheating if airflow is blocked, which may occur due to air circulation fan failure or overstocking of the cabinet. There are two MRST's in the TD- 1200 and TD-2400 located at the back of the cabinet.
For instructions on resetting the safety thermostat, refer to the manual.