Getting Started | Pharmacy Vaccine Refrigerators

Intended Use, Safety Guidelines, and Environmental Considerations

Download User Manual

Applicable Models:

TPR-145-1-GD
TPR-200-1-GD
TPR-520-1-GD
TPR-750-2-GD
TPR-950-2-GD
TPR-1500-3-GD

TPR-145-1-SD
TPR-200-1-SD
TPR-520-1-SD
TPR-750-2-SD
TPR-950-2-SD
TPR-1500-3-SD

📜 Quick Navigation

Intended Use

Unpacking

Moving

Operating Environment

Electrical Connections

BMS Output

 


Intended Use

The Thermoline Pharmacy Vaccine Refrigerator models 145, 200, 520, 950, and 1500 are designed and manufactured to maintain safe and precise temperature conditions for stored items. Operating at a fixed +5°C, with alarms set at 2°C and 8°C, these refrigerators provide industry-standard forced draught refrigeration.

  • Control Stability: +/- 0.5°C
  • Operating Temperature: fixed at +5°C

Uncrating/Unpacking

Unpacking process for foam wrapped or boxed

Unpacking Process
    • Unpacking Process 2In most cases, the refrigerator will be delivered foam-wrapped and on its castors via sensitive freight (Fig 1). The 145 and 200 models will be delivered in a box on a skid (Fig 2). 
    • The refrigerator may be delivered to remote areas in a crate. To remove the crate packaging, unscrew both the left and right sides. A forklift is required to remove the refrigerator from the crate.
    • Please only dispose of the packaging once the the refrigerator is inspected. If any damage is found, contact your supplier or Thermoline immediately at:

 

 

Moving

Moving the refrigerator on castors:

  • Ensure that the refrigerator is rolled on an even and flat surface. Uneven surfaces can cause the refrigerator to fall over. 

IMPORTANT

Refrigerators are 'Top Heavy'. Do not move the cabinet too quickly. (Fig 3 & 4)

 

Safe moving of cabinet   Unsafe moving of cabinet

 

 

Operating Environment

Refrigerator Location

  • Place the refrigerator away from direct sunlight and heat sources (e.g., heaters).
  • Ensure the room's ambient temperature does not exceed the refrigerator's operational limits.
  • Refrigerators should be stored inside at all times. Failure to adhere to this could cause significant drops in cabinet performance and damage to items stored inside.

Environmental Conditions

Condition Temperature Humidity
Extreme Operating Environment 10°C to 32°C Up to 85% RH
Optimal Environment 23°C (±2.0°C) 50% RH (±5% RH)

Bench Top LocationStacked Bench Top

  • Only the TPR-145-GD/SD model can be placed on a bench or plinth, as the other models are too large.
  • Under no circumstances should any refrigerators be stacked on top of each other (Fig. 2). 
  • Thermoline 145 refrigerators are equipped with adjustable feet in case of uneven surfaces. 

 

Electrical Connections

All refrigerators require a 10amp, 230V, 50Hz power supply

Important Notes

  • Always use a dedicated outlet for the refrigerator.
  • Do not use power boards or extension cords to connect the refrigerator.
 

Electrical Conditions

  • All refrigerators include a 2.5m removable mains power lead with a three-pin plug. 
  • Ensure the product is reasonably distanced from the power supply to prevent strain on the lead.

Warning Operating Environment Warnings

  • Refrigerators should be stored inside at all times. Failure to adhere to this could cause significant drops in cabinet performance and damage to items stored inside.  
  • Refrigerators are not suitable for use with flammable solvents! They are fitted with components that may be the source of ignition. 

 


BMS Output

Vaccine Fridges are fitted with a 3-pin socket for connection to a building monitoring system or phone dialler. A separate plug is provided to connect the socket to your system.

For detailed instructions on the BMS output, refer to the BMS Manual.