Getting Started | Humiditherm Cabinets with Lighting

Intended Use, Safety Guidelines, and Environmental Considerations

Download User Manual

Applicable Models:

TRHDL-300-SD
TRHDL-460-SD
TRHDL-850-SD

TRHSL-300-SD
TRHSL-460-SD
TRHSL-850-SD

📜 Quick Navigation

🔍 Need More Information?

📦 Setup & Installation

⚙️ Operating Procedures

🧰 Maintenance & Cleaning

🔧 Troubleshooting & Repair

📜 Warranty & Support

 


 

Intended Use

The Thermoline Humiditherm Cabinet with lighting is available in 300, 460, and 850 models, with either door lighting (DL) or shelf lighting (SL). Designed to operate within a temperature range of +5°C to +40°C with adjustable alarms, these cabinets provide industry-standard temperature and humidity control.

Humiditherm cabinets with lighting are only available as solid door models:

  • TRH-DL/SL models operate within a temperature range of +5°C to +40°C, with a maximum humidity of 90%.

Uncrating/Unpacking

Unpacking process for foam wrapped and crated:Unpacking ProcessCrate Unpacking Process

    • The Humiditherm cabinet will be delivered foam-wrapped and on its castors via sensitive freight (Fig 1) or in a crate.
    • If the cabinet is delivered on a crate, remove the packaging by unscrewing both the left and right sides. A forklift may be required to lift the cabinet off the crate safely.
    • Please only dispose of the packaging once the Humiditherm cabinet is inspected. If any damage is found, contact your supplier or Thermoline immediately at:
Phone: +61 2 9604 3911
Email: service@thermoline.com.au

 

Moving

Moving the Humiditherm cabinet:

  • Ensure that the Humiditherm cabinet is rolled on an even and flat surface. Uneven surfaces can cause the cabinet to fall over. 

IMPORTANT

 Humiditherm cabinets are 'Top Heavy'. Do not move the cabinet too quickly. (Fig 3 & 4)

Safe moving of Envirotherm cabinet        Unsafe moving of Envirotherm cabinet
 

Operating Environment

Cabinet Location

  • Place the cabinet away from direct sunlight and heat sources (e.g., heaters).
  • Ensure the room's ambient temperature does not exceed the cabinet's operational limits.
  • The Humiditherm cabinet should be stored inside at all times. Failure to adhere to this could cause significant drops in cabinet performance and damage to items stored inside.
  • If multiple cabinets are placed close together, heat buildup can occur due to airflow direction. One cabinet may transfer heat to the next, affecting performance. To prevent this, ensure cabinets are spaced more than 300mm apart or positioned so that hot air is not directed into adjacent units.

Thermoline Manual Image - Humiditherm and Climatron - Deflectors_b

Environmental Conditions

Condition Temperature Humidity
Extreme Operating Environment 10°C to 32°C Up to 85% RH
Optimal Environment 23°C (±2.0°C) 50% RH (±5% RH)

Electrical Connections

All Humiditherm cabinets require a 15amp, 230V, 50Hz power supply. Use a dedicated outlet for the cabinet, and avoid using power boards or extension cords. Each cabinet is supplied with a standard 3-pin moulded plug for connection to the mains.

Power Supply:

  • All Humiditherm cabinets come with a 2.5m removable mains power lead, featuring a three-pin plug and a female IEC plug. Ensure the cabinet is positioned at a reasonable distance from the power supply.

  • The cabinet itself has a male IEC socket.

Water Quality

For optimal performance, follow the recommended water quality parameters below. Due to the extensive use of stainless steel in Thermoline products, do not use deionised water. It can cause corrosion (due to leaching over time) that may not be covered under warranty.

 

Parameter Range
Resistivity 0.1 - 0.5 MΩ 
Conductivity 2-10 µs/cm
Total Dissolved Solids <10 mg/L
Acidity 6-7 pH

 

Avoid tap water if possible. Using tap water may significantly increase the required frequency of cleaning and maintenance of the equipment. Unacceptable water can cause excess scale build-up and mineral deposits, particularly in humidity systems. This, in turn, can cause heater failure and issues with oat switches. In water baths, corrosion due to insucient cleaning is the primary concern, with heater circulators and circulation pumps being the most signicant issue.

Note: All Thermoline equipment using water requires regular maintenance, inspection and cleaning. Six monthly for Humiditherm, Envirotherm and Climatron cabinets. Water baths will require much more frequent cleaning due to the ease of contamination in the water. Water should be changed for any signs of contamination. 

 

Warning WARNINGS

  • Humiditherm cabinets require adequate ventilation, with at least 300mm of clearance on either side and at the back. If multiple TEC or TRH cabinets are placed close together, additional spacing may be necessary to ensure proper airflow.
  • Humiditherm cabinets should be stored inside at all times. Failure to adhere to this could cause significant drops in cabinet performance and damage to items stored inside. 
  • Ensure that the feed water is suitable for the cabinet. 
  • Humiditherm cabinets are not suitable for use with flammable solvents! They are fitted with components that may be the source of ignition.