Troubleshooting and Repair | TD-AS1681 Drying Ovens

Identifying Common Issues, Quick Fixes, and When to Seek Professional Assistance

Applicable Models:

TD-150F-AS1681
TD-500F-AS1681
 

Need More Information?

📂 Getting Started

📦 Setup & Installation

⚙️ Operating Procedures

🧰 Maintenance & Cleaning

📜 Warranty & Support

 


Problem: Oven Temperature does not match the controller readout

  • Cause:

    1. Sensor Issue: The temperature sensor inside the oven may not be properly calibrated or functioning correctly. If the offset is greater than 5°C, the sensor may need replacing.
    2. Offset Issue: The controller's offset may have been inadvertently adjusted, causing the discrepancy.
  • Fix:

    1. Replace the oven sensor if necessary. Contact a technician to assist with replacement and calibration.
    2. Check and adjust the offset in the controller by selecting "iNS" (Input Shift) and following the calibration instructions.

Part Number: 40716 - PT100 Sensor


Problem: The "ALM" light is illuminated on the controller and the PV is reading far higher than usual

  • Cause:
    The temperature offset has likely been inadvertently adjusted.

  • Fix:
    Follow the sensor calibration instructions to reset the "iNS" (Input Shift) value back to zero.


There is no air flow in the oven

  • Cause:
    The internal fan may have failed, limiting air circulation within the oven.

  • Fix:
    Replace the internal fan to restore air circulation.

Part Number: 50429 - D Series Motor


Problem: The word "Stop" is showing on the controller

  • Cause:
    The controller is in the "STOP" mode, preventing normal operation.

  • Fix:

    1. Press the "PAGE" and "SCROLL" buttons simultaneously until "oAPt" appears on the screen.
    2. Press the "SCROLL" button to select the parameter "PMSK."
    3. Press "DOWN" to turn it off.
    4. Navigate back to the main menu using the "PAGE" and "SCROLL" buttons.
    5. Press "SCROLL" until "R-S" appears on the screen.
    6. Use the "DOWN" button to change "STOP" to "RUN."
    7. Follow these steps again to ensure "PMSK" is turned back on.

    Note: Ensure no other parameters are changed during this process to avoid additional issues.


Problem: Purge light out (TD-AS1681 only)

  • Cause:
    The pressure differential has tripped due to abrupt door closing, blockage in the inlet or outlet port, or failure of the fan motors.

  • Fix:

    1. When closing the doors, do so gently to avoid creating an unnecessary pressure differential.
    2. Inspect the inlet and outlet ports for any blockages and clear them as needed.
    3. Check the fan motors and replace them if they have failed.

Part Number: 50429 - D Series Motor


📞 Contacting Thermoline for Product Information and Support

When reaching out to Thermoline for product information, please have the Serial Number and any other relevant details on hand. The serial number can be found on a white sticker, typically located near the power IEC socket.

For assistance, contact Thermoline Service:

Having this information ready helps us provide efficient and accurate support.