Troubleshooting and Repair | Refrigerated Incubator TMLR-200

Identifying Common Issues, Quick Fixes, and When to Seek Professional Assistance

Applicable Models:

TMLR-200  

🔍 Need More Information?

📂 Getting Started

📦 Setup & Installation

⚙️ Operating Procedures

🧰 Maintenance & Cleaning

📜 Warranty & Support

 


Problem: Battery Alarm still occurs every 7 days, even after being acknowledged and reset.

  • Cause:
    The backup battery's voltage may be low, indicating the need for a replacement.
  • Fix:
    Replace the backup battery with the following part:
    Part Number: 70320 - Sealed Lead Acid Battery Backup (12V 9.5A).

Problem: Unable to acknowledge the alarm.

  • Cause:
    The chamber may still be in an alarm condition. Alarms cannot be acknowledged until the underlying issue is resolved.
  • Fix:
  1. Confirm that the chamber is no longer in an alarm condition. Resolve any issues causing the alarm.
  2. Muting the alarm will temporarily silence it for 15 minutes but will not clear the condition.
  3. After resolving the issue, press and hold the alarm acknowledgment button for 10 seconds to fully acknowledge and clear the alarm.

📞 Contacting Thermoline for Product Information and Support

When reaching out to Thermoline for product information, please have the Serial Number and any other relevant details on hand. The serial number can be found on a white sticker, typically located near the power IEC socket.