Troubleshooting and Repair | Platelet Shaker
Identifying Common Issues, Quick Fixes, and When to Seek Professional Assistance
Applicable Models:
TPS-18 |
🔍 Need More Information?
📂 Getting Started
📦 Setup & Installation
⚙️ Operating Procedures
🧰 Maintenance & Cleaning
📜 Warranty & Support
Problem: Shaker won't start
Cause:
- Not plugged into power correctly or faulty power point
Fix:
- Check the power connection and supply.
Problem: Excessive noise
Cause:
- External or internal rubbing of the carriage
Fix:
- Check for clearance around the carriage.
Problem: Excessive heat in the base of the case.
Cause:
- Fan motor blocked or broken.
Fix:
- Ensure vents are clear.
- Check the fan motor.
Problem: Power is turned on but the carriage does not move.
Cause:
- Connecting rod damaged.
- Faulty motor.
Fix:
- Inspect the connecting rod.
- Check the motor.
📞 Contacting Thermoline for Product Information and Support
When reaching out to Thermoline for product information, please have the Serial Number and any other relevant details on hand. The serial number can be found on a white sticker, typically located near the power IEC socket.
For assistance, contact Thermoline Service:
- Phone: +61 2 9604 3911
- Email: service@thermoline.com.au
Having this information ready helps us provide efficient and accurate support.