Problem: Shaker does not power up even when switched on
Cause:
- The power cable is not connected.
- The fuse is broken.
Fix:
- Ensure the power cable is securely connected to both the shaker and the power outlet.
- Check the fuse and replace it if necessary.
Problem: Error Code 'Er 01' is showing.
Cause:
- Motor blocked
- Press the speed control knob or timer control knob to stop the shaking.
- Restart the shaking with a reduced load.
- If shaking does not start, further reduce the load and try again.
Problem: Error Code 'Er 02' is showing
Cause:
- Motor overspeed
- Press the speed control knob or timer control knob to stop the shaking.
- Restart the shaking with a reduced load.
- If shaking does not start, further reduce the load and try again.