Troubleshooting and Repair | Laboratory Refrigerators
Identifying Common Issues, Quick Fixes, and When to Seek Professional Assistance
Applicable Models:
TLR-145-1-GD TLR-200-1-GD TLR-520-1-GD TLR-750-2-GD TLR-950-2-GD TLR-1500-3-GD |
TLR-145-1-SD |
🔍 Need More Information?
📂 Getting Started
📦 Setup & Installation
⚙️ Operating Procedures
🧰 Maintenance & Cleaning
📜 Warranty & Support
Problem: Temperature of the freezer is running warmer than expected.
Causes:
- The condenser may be dirty or blocked.
- Recently added items may be affecting the internal temperature.
- Air flow may be blocked.
- Incorrect offset.
Fix:
- Clean condenser as described within manual.
- If new samples have been added, allow about an hour to see if the temperature stabilises and comes down.
- Ensure the stock is evenly distributed.
- Air temperature offset may require to be adjusted.
Problem: PoF is shown on the Dixell Controller.
Cause:
- The message "PoF" indicates that the controller is locked.
Fix:
- Press the UP and DOWN buttons simultaneously until PoN appears.
Problem: Condensation on the outside of the door.
Cause:
- Ambient Humidity is too high.
Fix:
- Reduce ambient humidity by improving ventilation or using an air conditioner or dehumidifier in the room.