Troubleshooting and Repair | Economy Laboratory Freezers
Identifying Common Issues, Quick Fixes, and When to Seek Professional Assistance
Applicable Models:
TF-160-1-SD TF-260-1-SD |
🔍 Need More Information?
📂 Getting Started
📦 Setup & Installation
⚙️ Operating Procedures
🧰 Maintenance & Cleaning
📜 Warranty & Support
Problem: Temperature of the freezer is running hotter than expected.
Causes:
- The condenser may be dirty or blocked.
- Recently added items may be affecting the internal temperature.
Fix:
- Clean condenser as described within manual.
- If new samples have been added, allow about an hour to see if the temperature stabilises and comes down.
Problem: PoF is shown on the Dixell Controller.
Cause:
- The message "PoF" indicates that the controller is locked.
Fix:
- Press the UP and DOWN buttons simultaneously until PoN appears.
Problem: The buttons on the controller are damaged, and the controller does not respond, even though the display is working.
Cause:
- Faceplate is damaged.
Fix:
- Replace the faceplate.