Min/Max Reading on a Pharmacy Fridge - Dixell Controller

Reading the Minimum, Maximum temperature and Resetting a Dixell Controller

MicrosoftTeams-image (14)
HOW TO: SEE THE MINIMUM TEMPERATURE
1. Press and release the DOWN key.
2. The “Lo” message will be displayed followed by the minimum temperature recorded.
3. By pressing the DOWN key again or by waiting 5 seconds the normal display will be restored.
HOW TO: SEE THE MAXIMUM TEMPERATURE
1. Press and release the UP key.
2. The “Hi” message will be displayed followed by the maximum temperature recorded.
3. By pressing the UP key again or by waiting 5 seconds the normal display will be restored.
HOW TO: RESET THE MAXIMUM AND MINIMUM TEMPERATURE
1. Keep the SET key pressed more than 3 seconds, while the max or min temperature is displayed. (“rSt” message will be displayed)
2. To confirm the operation the “rSt” message will start blinking and the normal temperature will be displayed.