Getting Started | Premium Incubators

Intended Use, Safety Guidelines, and Environmental Considerations

Download User Manual

Applicable Models:

TI-520F-SD
TI-520F-GD
TI-1100F-SD
TI-1100F-GD

 

📜 Quick Navigation

🔍 Need More Information?

📦 Setup & Installation

⚙️ Operating Procedures

🧰 Maintenance & Cleaning

🔧 Troubleshooting & Repair

📜 Warranty & Support

 


Intended Use

The Thermoline Laboratory Premium Incubators are designed and manufactured to provide a controlled and precise temperature environment for stored items. These incubators operate within a temperature range of Ambient +5°C to +60°C, with a control accuracy of ±0.1°C.

To ensure optimal performance, the incubator must be used indoors and should not be placed in environments where the ambient temperature exceeds 32°C.

Key Features:

  • Operating temperature from ambient +5°C to +60°C
  • Ecofoam Insulation
  • Electronically controlled heating element
  • Monitoring port hole
  • Removable and adjustable plastic coated shelves
  • Optional glass or solid doors

Unpacking

Unpacking ProcessIn most cases, the incubator will be foam-wrapped and delivered on its castors via sensitive freight. (Fig 1)

For remote deliveries, the incubator may arrive in a crate. To remove the crate packaging:

  • Unscrew both the left and right sides of the packaging. (Fig 2)
  • A forklift is required to safely remove the incubator from the crate.

Unpacking Process of IncubatorDo not dispose of the packaging until the incubator has been fully inspected. If any damage is detected upon opening, contact your supplier or Thermoline immediately on +61 2 9604 3911 or email at service@thermoline.com.au.

 

Moving and Lifting

Moving the incubator:

  • Ensure that the incubator is rolled on an even and flat surface. Uneven surfaces can cause the cabinet to fall over. 

IMPORTANT

 Incubators are 'Top Heavy'. Do not move the cabinet too quickly. (Fig 3 & 4)

        Safe Moving of Incubator  Unsafe Moving of Incubator

Operating Environment

Location

  • Place the incubator away from direct sunlight and heat sources (e.g., heaters).
  • Ensure the room's ambient temperature does not exceed the cabinet's operational limits.
  • The Incubator should be stored inside at all times. Failure to adhere to this could cause significant drops in cabinet performance and damage to items stored inside.

 Environmental Conditions

Condition Temperature Humidity
Extreme Operating Environment 10°C to 32°C Up to 85% RH
Optimal Environment 23°C (±5°C) 50% RH (±25% RH)

Electrical Connections

All incubators require a 10amp, 230V, 50Hz power supply. Use a dedicated outlet for the cabinet, and avoid using power boards or extension cords. Each cabinet is supplied with a standard 3-pin moulded plug for connection to the mains.

Power Supply:

  • All incubators come with a 2.5m removable mains power lead, featuring a three-pin plug and a female IEC plug. Ensure the cabinet is positioned at a reasonable distance from the power supply.

  • The cabinet itself has a male IEC socket.

 

Warning WARNINGS

  • The incubator should be stored indoors at all times. Failure to adhere to this could cause significant drops in cabinet performance and damage to items stored inside.

  • Incubators are not suitable for use with flammable solvents, as they are fitted with components that may be a source of ignition.

  • The incubator will generate heat during use. As a result, the inner liner of the cabinet may be hot to touch. Proceed with caution.